HIV Test Kit Procedure

Leave a message

*Preparation Before the Experiment
1. Remove the test kit and the samples to be tested. Allow all reagents and samples to reach room temperature (18–25°C).
2. Adjust the incubator or water bath to the required reaction temperature.
3. Dilute the wash buffer 20-fold with distilled water and set aside.


*Experimental Procedure
1. Take the required number of coated wells and label them. Designate one well as a blank control. Add 50 μl each of the negative control (in triplicate) and positive control (in duplicate), as well as 50 μl of each sample to be tested, into their respective reaction wells.
2. Apply the plate sealer.
3. Remove the plate sealer and discard the liquid from the wells. Wash the wells 6 times, then blot them dry on absorbent paper.
4. Add 50 μl of the enzyme conjugate to each well (do not add to the blank control well), and re-apply the plate sealer.
5. Incubate in a 37°C incubator for 30 minutes.
6. Remove the plate sealer, discard the liquid from the wells, wash the wells 6 times, and blot them dry on absorbent paper.
7. Add 50 μl each of the substrate solution and the chromogenic agent to every well. Incubate at 37°C in the dark for 10 minutes.
8. Add 50 μl of the stop solution to each well, gently shake to mix, and then immediately proceed to read the results.
9. Use a microplate reader for measurement. For single-wavelength measurement, zero the instrument using the blank well and measure the optical density (OD) of each well at 450 nm. Alternatively, use a dual-wavelength measurement (450 nm with a reference wavelength of 620–630 nm) to determine the OD value for each well.